FDA Brief

Company record

Mint Medical GmbH

0 FDA warning letters since 2023 and 1 recall in the last 12 months.

Get an email the day FDA posts anything new about Mint Medical GmbH.

Recalls

ReportedClassFirmProduct and reason
May 13 Class II Mint Medical GmbH
Device
Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.;
If the connection between a mint Lesion workstation and the mint Lesion server is interrupted while performing a read, in rare cases some information may be lost or incorrectly linked to other patients.